Cleaning the Trough
Put on gloves and fill a 1 liter
bottle with pure water from the Milli-Q
purifier.
Remove cover of the trough and the
barrier. To clean the barrier, please see below.
Pour water into the trough until
it is just full, i.e. not bulging over.
Add Hellmanex
soap, 2 milliliters is a good amount.
If the trough is temperature
regulated, turn the thermostat up to about 40 degrees Celsius.
Let the water sit for at least 30
minutes.
Empty and rinse the trough 6 times to make sure the Hellmanex
is gone and the trough is clean.
Cleaning the barrier:
The barrier should first be
cleaned with ethanol and rinsed with pure water.
Place the barrier in a rather
large, clean, plastic container filled with pure
water from the Milli-Q purifier.
Add Hellmanex
soap, about 2 milliliters, to the water.
Place this container in the
sonicator, which should be filled with pure water also. Set it for
15 minutes.
After it is done, rinse out the
container and barrier with pure water.
Fill the container again with pure
water and insert the barrier again.
Put the container with the barrier
in it in the sonicator again for another 15 minutes.
Once it is done, rinse out the
container and barrier again.
The barrier is now clean, you are
ready to insert it into a clean trough.
|
|